FDA recall Z-0187-2019

Alcon Research, LTD. · Class I · device

Product

CyPass ULTRA SYSTEM, CYPASS MICRO-STENT, CYPASS APPLIER, REF CATALOG #: 8065754007. For ophthalmic use.

Reason for recall

Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.

Key facts

Status
Terminated
Initiation date
2018-08-28
Report date
2018-10-31
Termination date
2020-12-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sinking Spring, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0187-2019