FDA recall Z-0187-2020

Life Technologies Holdings Pte Ltd · Class II · device

Product

Torrent Suite Dx Software Version 5.6.4 Catalog Number:A33178 OEM Customer software: Sentosa SQ Suite Software v5.6.20 UDI: (01)10190302005654 (11)000000 (10)5.6.4 (240)A33178

Reason for recall

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Distribution

states : AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA OUS: South Korea, ITALY, China, Japan, Singapore, Germany, TAIWAN, Poland

Key facts

Status
Terminated
Initiation date
2019-09-26
Report date
2019-10-30
Termination date
2021-11-23
Voluntary/Mandated
Voluntary: Firm initiated