FDA recall Z-0189-2018

Volcano Corporation · Class II · device

Product

Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF 400-0100.01; CORE Mobile Imaging System Refurbished REF 400-0100.01-R; CORE Mobile Imaging System (240V) REF 400-0100.07; CORE Mobile Imaging System Refurbished REF 400-0100.07-R; CORE Mobile Imaging System (100V) REF 400-0100.08; CORE Mobile Imaging System Refurbished REF 400-0100.08-R; CORE Imaging System REF 400-0100.02 For intravascular ultrasound imaging.

Reason for recall

On certain Impacted Systems, an unexpected Microsoft Windows Security dialog may appear during use, and the users response to the dialog may adversely affect the subsequent operation of the device.

Distribution

Worldwide distribution including US nationwide and Canada, AE, AT, AU, BD, BE, BG, BR, CH, CL, CN, CY, DE, DK, EG, ES, FR, GB, HU, IE, IL, IN, IT, JO, JP, KW, KR, LB, MA, MX, MY, NL, NO, PA, PK, PL, PT, QA, RO, RU, SA, SE, SG, SK, Sri Lanka, TH, TW, UAE, UK, ZA,

Key facts

Status
Terminated
Initiation date
2017-11-03
Report date
2017-12-06
Termination date
2023-10-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rancho Cordova, CA, United States