FDA recall Z-0192-2022

FujiFilm Healthcare Americas Corporation · Class II · device

Product

Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0

Reason for recall

When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2021-10-14
Report date
2021-11-10
Voluntary/Mandated
Voluntary: Firm initiated