FDA recall Z-0193-2018

Pro-Med Instruments Gmbh · Class II · device

Product

pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.

Reason for recall

The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.

Distribution

Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada

Key facts

Status
Terminated
Initiation date
2017-11-02
Report date
2017-12-13
Termination date
2018-07-11