FDA recall Z-0200-2022

Siemens Medical Solutions USA, Inc · Class II · device

Product

Artis zeego, Fluoroscopic X-Ray System Models: 10280959,10502505, 10848283 in combination with the following workplace/system application software models: Product/Model #: syngo X Workplace, 10281061 syngo X Workplace, 10281299 syngo X Workplace, 10502522 Syngo Application Software, 10848815

Reason for recall

Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5 degrees relative to the patient and to be used inappropriately by the system and result in incorrect treatment of the patient

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2021-09-24
Report date
2021-11-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States