FDA recall Z-0202-2020

Ethicon Endo-Surgery Inc · Class I · device

Product

ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A

Reason for recall

The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

Distribution

Worldwide distribution. US Nationwide. Belgium, Brazil, Canada, Chile, Colombia, Japan, Kuwait, Saudi Arabia, and Singapore.

Key facts

Status
Terminated
Initiation date
2019-10-03
Report date
2019-11-06
Termination date
2021-06-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Blue Ash, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0202-2020