FDA recall Z-0213-2022

Philips North America Llc · Class II · device

Product

Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)

Reason for recall

When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.

Distribution

Domestic distribution nationwide. International distribution worldwide.

Key facts

Status
Terminated
Initiation date
2021-09-24
Report date
2021-11-17
Termination date
2024-09-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0213-2022