# FDA recall Z-0213-2026

> **Wipro GE Healthcare Private Ltd.** · Class II · device recall initiated 2025-09-05.

## Product

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

## Reason for recall

GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

## Distribution

Worldwide distribution.

## Key facts

- **Recall number:** Z-0213-2026
- **Recalling firm:** Wipro GE Healthcare Private Ltd.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-09-05
- **Report date:** 2025-10-29

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Bengaluru, India

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0213-2026

## Citation

> AI Analytics. FDA recall Z-0213-2026. Retrieved 2026-09-14 from https://api.ai-analytics.org/recall/Z-0213-2026. Source: US FDA. Licensed CC0.

---