FDA recall Z-0216-2026

Beckman Coulter, Inc. · Class II · device

Product

DxI 9000 Access Immunoassay Analyzer, Part Number C11137

Reason for recall

Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-09-19
Report date
2025-10-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Chaska, MN, United States