# FDA recall Z-0222-2025

> **Smiths Medical ASD Inc.** · Class I · device recall initiated 2024-07-16.

## Product

CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems:  21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-

## Reason for recall

Damage to the battery pack may have caused a short to a capacitor within the battery pack.  While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case.  If this issue occurs, the battery pack charging circuit may become inoperable.

## Distribution

Worldwide Distribution

## Key facts

- **Recall number:** Z-0222-2025
- **Recalling firm:** Smiths Medical ASD Inc.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-07-16
- **Report date:** 2024-11-06

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Minneapolis, MN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0222-2025

## Citation

> AI Analytics. FDA recall Z-0222-2025. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-0222-2025. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*