FDA recall Z-0226-2022

Siemens Medical Solutions USA, Inc · Class II · device

Product

Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies

Reason for recall

System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs

Distribution

Nationwide

Key facts

Status
Ongoing
Initiation date
2021-10-05
Report date
2021-11-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States