FDA recall Z-0226-2023

Boston Scientific Corporation · Class II · device

Product

ORISE ProKnife Procedure Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device. UPN (1)ORISE ProKnife 1.5 mm Electrode - Kit M00519380; (2) ORISE ProKnife 2.0 mm Electrode - Kit M00519390; (3) ORISE ProKnife 3.0 mm Electrode - Kit M00519400

Reason for recall

***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention

Distribution

Worldwide Distribution: US (Nationwide) and (Foreign) countries of: Algeria Argentina Argentina Australia Austria Bahrain Belgium Brazil Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Iraq Ireland Israel Italy Jordan Kuwait Latvia Lebanon Lithuania Luxembourg Mexico Netherlands New Zealand Norway Oman Palestine Panama Per¿ Poland Portugal Romania Russian Fed. Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Unit.Arab Emir. Zimbabwe

Key facts

Status
Ongoing
Initiation date
2022-10-10
Report date
2022-11-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marlborough, MA, United States