FDA recall Z-0230-2026

CPM Medical Consultants, LLC. · Class II · device

Product

Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS421208, OS421210, OS421212, OS421000L, OS421000R, OS421002L, OS421002R, OS421004L, OS421004R, OS421006L, OS421006R, OS42260LL, OS42260LR, OS42260ML, OS42260MR, OS42260SL, OS42260SR, OS42260XSL, OS42260XSR, OS422800, OS422803, OS422804, OS422805, OS422807, OS422825, OS422400, OS422403, OS422404, OS422405, OS422406, OS422407, OS421406, OS421408, OS421414, OS422702, OS422703, OS422712, OS422714, OS422716, OS421512, OS421514, OS421516, OS421112, OS421116, OS421120, OS421124, OS421130, OS42150L, OS42150M, OS42150S, OS42150XS, OS326920-T, OS326922-T, OS326924-T, OS326925, OS326925-T, OS326926-T, OS

Reason for recall

Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.

Distribution

US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.

Key facts

Status
Ongoing
Initiation date
2025-09-17
Report date
2025-10-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Farmers Branch, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0230-2026