FDA recall Z-0231-2023

Baxter Healthcare Corporation · Class II · device

Product

Hill-Rom Progressa Bed System

Reason for recall

The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI, IN, LA, MA, MO, NC, NM, NV, NY, OK, OR, PA, SC. TN, TX, WA, and WI. (UPDATED: States of PA and SC and there was government distribution.) There was no military distribution. Foreign distribution was made to Canada, Qatar, Singapore, and UAE. UPDATE: Saudi Arabia now included.

Key facts

Status
Ongoing
Initiation date
2022-10-18
Report date
2022-11-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0231-2023