FDA recall Z-0232-2022

Arrow International Inc · Class I · device

Product

Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-HFWC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venousfistulae and synthetic dialysis grafts

Reason for recall

7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.

Distribution

Worldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: BE CA CL CZ DE ES FR IL IT KR PH SA TR TW UK

Key facts

Status
Terminated
Initiation date
2021-09-20
Report date
2021-11-24
Termination date
2023-10-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Reading, PA, United States