FDA recall Z-0232-2023

Becton Dickinson & Co. · Class II · device

Product

BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607

Reason for recall

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2022-10-13
Report date
2022-11-23
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sparks, MD, United States