FDA recall Z-0232-2025

Olympus Corporation of the Americas · Class II · device

Product

Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.

Reason for recall

Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention

Distribution

OH

Key facts

Status
Ongoing
Initiation date
2024-09-24
Report date
2024-11-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Center Valley, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0232-2025