FDA recall Z-0235-2019

Stryker Medical Division of Stryker Corporation · Class II · device

Product

Altrix¿ Precision Temperature Management System Model: 8001Product Usage: The Altrix system is intended for circulating temperature controlled warm or cold water via patient contact thermal transfer devices for the application of regulating human body temperature in situations where a physician or clinician with prescription privileges determines that temperature therapy is necessary or desirable.

Reason for recall

Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5¿C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical interference between the device fan and filter

Distribution

Worldwide - US Nationwide Distribution and the countries of Foreign: Australia, Canada, Netherlands, Switzerland and United Kingdom

Key facts

Status
Terminated
Initiation date
2018-09-06
Report date
2018-10-31
Termination date
2020-04-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States

Primary source