FDA recall Z-0257-2025

Covidien LP · Class II · device

Product

Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm x 75cm (CFN CH07-40-75US), 8mm x 40mm x 75cm (CFN CH08-40-75US), 9mm x 40mm x 75cm (CFN CH09-40-75US), 10mm x 40mm x 75cm (CFN CH10-40-75US), 12mm x 40mm x 75cm (CFN CH12-40-75US)

Reason for recall

Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Canada, Canary Islands, China, France, French Guiana, Germany, Greece, Guadeloupe, India, Ireland, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Spain and Turkey.

Key facts

Status
Ongoing
Initiation date
2024-09-19
Report date
2024-11-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mansfield, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0257-2025