FDA recall Z-0261-2018

Orthoscan, Inc. · Class II · device

Product

Mobile Mini C-arm system Part # 1000-0004-FD. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.

Reason for recall

OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-08-31
Report date
2018-01-17
Termination date
2021-04-21
Voluntary/Mandated
FDA Mandated
Location
Scottsdale, AZ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0261-2018