FDA recall Z-0261-2025

Ambu Inc. · Class II · device

Product

Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)

Reason for recall

Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

Distribution

US Nationwide. Canada.

Key facts

Status
Ongoing
Initiation date
2024-09-16
Report date
2024-11-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Columbia, MD, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0261-2025