L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.

## Reason for recall

L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel.  This a gap in x-ray shielding  can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).

## Distribution

USA (nationwide)  Distribution.  Owned and operated by the U.S. Transportation Security Administration. Location of individual products is law enforcement sensitive. TSA HQ is located in Crystal City, Virginia.

## Key facts

- **Recall number:** Z-0270-2018
- **Recalling firm:** L3 Security & Detection
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-10-27
- **Report date:** 2018-01-03
- **Termination date:** 2020-02-10

- **Voluntary/Mandated:** FDA Mandated
- **Location:** Woburn, MA, United States