FDA recall Z-0273-2018

Ortho-Clinical Diagnostics · Class II · device

Product

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.

Distribution

Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.

Key facts

Status
Terminated
Initiation date
2017-06-12
Report date
2018-01-03
Termination date
2020-08-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0273-2018