FDA recall Z-0276-2022

Instrumentation Laboratory · Class II · device

Product

HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

Reason for recall

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion includes states of IN, MN, and NC.

Key facts

Status
Ongoing
Initiation date
2021-10-13
Report date
2021-12-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bedford, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0276-2022