FDA recall Z-0281-2019

Leica Microsystems, Inc. · Class II · device

Product

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Reason for recall

The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.

Distribution

Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.

Key facts

Status
Terminated
Initiation date
2018-09-17
Report date
2018-11-07
Termination date
2020-08-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Buffalo Grove, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0281-2019