FDA recall Z-0291-2018

Orthofix Srl · Class II · device

Product

Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini Kit Instruments Sterile, Galaxy UNYCO Diaphyseal Femur Box, Galaxy UNYCO Knee Bridging Box, Galaxy UNYCO Ankle Bridging Box - Unilateral Frame, Galaxy UNYCO Ankle Bridging Box - Delta Frame, and Galaxy UNYCO Diaphyseal Tibia Box. Intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

Reason for recall

The power drill torque limiter may break during use in surgery application.

Distribution

Worldwide Distribution -- US, including the states of CA, FL, & PA; and, the countries of Australia, Germany, Finland, France, Italy, Lithuania, Netherlands, Singapore, South Africa, Sweden, Switzerland, & United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-07-05
Report date
2018-01-03
Termination date
2023-01-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bussolengo, N/A, Italy

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0291-2018