FDA recall Z-0297-2019

Siemens Medical Solutions USA, Inc · Class II · device

Product

Multix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with Tube stand 3D V (model number 7042042 or 7042059); Multix Fusion Digital Portable (model number 10746666) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Integrated (model number 10746667) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Wireless (model number 10893300) with Tube stand 3D V (model number 7042042)

Reason for recall

Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning

Distribution

US nationwide distribution

Key facts

Status
Terminated
Initiation date
2018-08-03
Report date
2018-11-07
Termination date
2019-11-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0297-2019