FDA recall Z-0299-2019

Siemens Medical Solutions USA, Inc · Class II · device

Product

Ysio (model number 10281013 or 10281163) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used for radiographic and tomographic exposures of the whole body.

Reason for recall

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Distribution

The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.

Key facts

Status
Terminated
Initiation date
2018-08-06
Report date
2018-11-07
Termination date
2020-04-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source