FDA recall Z-0309-2018

Beckman Coulter Inc. · Class III · device

Product

Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reason for recall

Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.

Distribution

Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.

Key facts

Status
Terminated
Initiation date
2017-11-06
Report date
2018-01-10
Termination date
2019-11-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States

Primary source