FDA recall Z-0316-2025

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION · Class II · device

Product

SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic examinations, including general radiography and pediatric examinations, excluding mammography.

Reason for recall

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Distribution

US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

Key facts

Status
Ongoing
Initiation date
2024-09-06
Report date
2024-11-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Kyoto, N/A, Japan

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0316-2025