# FDA recall Z-0322-2018

> **CareFusion 303, Inc.** · Class II · device recall initiated 2017-11-17.

## Product

Alaris Syringe Pump, Model 8110.  The firm name on the label is CareFusion.    The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous, subcutaneous or epidural. The Alaris Syringe pump and PCA modules can be connected to the Alaris PC unit which is the central programming, monitoring, and power supply component for the Alaris System. The syringe pump and PCA pump uses standard, plastic piston syringes and single-use administration sets with luer-lock connectors.

## Reason for recall

The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.

## Distribution

Worldwide Distribution - USA (nationwide) Distribution. There was also government and military distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahrain, Canada, Switzerland, United Kingdom, Kuwait, Mexico, Malaysia, The Netherlands, New Zealand, Qatar, Saudi Arabia, Taiwan, and South Africa.

## Key facts

- **Recall number:** Z-0322-2018
- **Recalling firm:** CareFusion 303, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-11-17
- **Report date:** 2018-01-17
- **Termination date:** 2020-05-04

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Diego, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0322-2018

## Citation

> AI Analytics. FDA recall Z-0322-2018. Retrieved 2026-06-25 from https://api.ai-analytics.org/recall/Z-0322-2018. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*