FDA recall Z-0331-2018

Philips Medical Systems (Cleveland) Inc · Class II · device

Product

Philips Healthcare Brilliance iCT computed tomography x-ray system

Reason for recall

There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.

Distribution

107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts

Key facts

Status
Terminated
Initiation date
2017-11-08
Report date
2018-01-17
Termination date
2020-04-17
Voluntary/Mandated
Voluntary: Firm initiated