FDA recall Z-0331-2023

Remel Inc · Class I · device

Product

Thermo SCIENTIFIC, Sensititre CMC5VGNF, Gram Negative IVD AST

Reason for recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Distribution

US Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2022-10-20
Report date
2022-12-14
Termination date
2024-09-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lenexa, KS, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0331-2023