FDA recall Z-0334-2024

Aizu Olympus Co., Ltd. · Class II · device

Product

Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.

Reason for recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2023-09-25
Report date
2023-11-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Aizuwakamatsu, N/A, Japan

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0334-2024