FDA recall Z-0345-2020

Abiomed, Inc. · Class II · device

Product

lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI

Reason for recall

This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.

Distribution

The product was distributed to GA

Key facts

Status
Terminated
Initiation date
2019-09-04
Report date
2019-11-13
Termination date
2020-01-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Danvers, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0345-2020