FDA recall Z-0349-2023

Primus Corporation dba Trinity Biotech · Class III · device

Product

A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis

Reason for recall

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Distribution

Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.

Key facts

Status
Terminated
Initiation date
2021-10-05
Report date
2022-12-07
Termination date
2023-08-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Kansas City, MO, United States