FDA recall Z-0351-2018

Hologic, Inc. · Class II · device

Product

Affirm Lateral Arm Upright Biopsy Accessory. Product Number: ASY-09880

Reason for recall

The firm is implementing a product correction involving its Affirm¿ Lateral Arm Upright Biopsy Accessory, ASY-09880. The correction has been initiated due to a component (the Left-hand Guide or the Blue Needle Guide) issue that could impact needle alignment for all the left-approach lateral biopsy procedures. This may have an impact on the position of the needle within the breast, and thus on the aspiration of the intended tissue. This impact is case-dependent and detectable when imagining is used to confirm needle position and/or tissue acquisition during and after a procedure. There are no known instances of illness or injury.

Distribution

US and OUS

Key facts

Status
Terminated
Initiation date
2017-07-07
Report date
2018-01-24
Termination date
2018-04-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Danbury, CT, United States

Primary source