FDA recall Z-0355-2024

B. Braun Medical, Inc. · Class I · device

Product

Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; 362911 Priming volume: 14mL, length: 84in, CARESITE, Spin-Lock; 363010 Priming volume: 21mL, length: 121in, SafeLine; 363019 Priming volume: 40mL, length: 123in, SafeLine; 363030 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 363032 20 drops/mL, priming volume: 20mL, length: 119in, SafeLine; 363230 15 drops/mL, priming volume: 22mL, lengt

Reason for recall

In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

Distribution

Distribution US nationwide, Canada and Germany.

Key facts

Status
Ongoing
Initiation date
2023-11-01
Report date
2023-12-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Allentown, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0355-2024