FDA recall Z-0357-2024

Instrumentation Laboratory · Class II · device

Product

GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004

Reason for recall

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Distribution

US States: MA, MN, NY. India.

Key facts

Status
Ongoing
Initiation date
2023-09-29
Report date
2023-11-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bedford, MA, United States