FDA recall Z-0365-2024

SunMed Holdings, LLC · Class II · device

Product

Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflation Bag Reservoir, Manometer, PEEP Valve, and 40cm Pop-Off REF # PRO-1927.

Reason for recall

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Distribution

US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.

Key facts

Status
Ongoing
Initiation date
2023-10-11
Report date
2023-11-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grand Rapids, MI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0365-2024