# FDA recall Z-0366-2023

> **Full Vision Inc** · Class II · device recall initiated 2022-08-31.

## Product

FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928

## Reason for recall

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

## Distribution

Wi, GA, NY, Mexico

## Key facts

- **Recall number:** Z-0366-2023
- **Recalling firm:** Full Vision Inc
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2022-08-31
- **Report date:** 2022-12-14

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Newton, KS, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0366-2023

## Citation

> AI Analytics. FDA recall Z-0366-2023. Retrieved 2026-05-27 from https://api.ai-analytics.org/recall/Z-0366-2023. Source: US FDA. Licensed CC0.

---