FDA recall Z-0367-2023

Full Vision Inc · Class II · device

Product

FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929

Reason for recall

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Distribution

Wi, GA, NY, Mexico

Key facts

Status
Ongoing
Initiation date
2022-08-31
Report date
2022-12-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Newton, KS, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0367-2023