FDA recall Z-0375-2023

Carefusion 2200 Inc · Class II · device

Product

Carefusion V. Mueller" SCOVILLE-GREENWOOD BAYONET BIPOLAR FORCEPS 1.5MM TIP INSULATED, OVERALL LENGTH 7-3/4" (195MM), REF F-1003

Reason for recall

BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.

Distribution

Nationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand

Key facts

Status
Ongoing
Initiation date
2022-11-02
Report date
2022-12-14
Voluntary/Mandated
Voluntary: Firm initiated