FDA recall Z-0379-2018

Carl Zeiss Metrology Inc · Class II · device

Product

Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.

Reason for recall

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Distribution

Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2017-11-14
Report date
2018-02-07
Voluntary/Mandated
FDA Mandated
Location
Maple Grove, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0379-2018