FDA recall Z-0385-2019

CareFusion 303, Inc. · Class II · device

Product

Alaris Infusion Pump administration sets labeled as: -Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 10011301, CareFusion, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP FITMENT ON PUMP SEGMENT IS OPEN, Fluid path is STERILE and NONPYROGENIC. 12276678 Rev 00 -VersaSafe Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump, REF 2120-0500, WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-01104 Rev 00 -SmartSite Infusion Set Alaris Pump Module/Medley Pump Module/Gemini Infusion Pump REF 2420-0500,WARNING: TO PREVENT FREE-FLOW, CLOSE SET CLAMP WHEN SAFETY CLAMP ON PUMP SEGMENT IS OPEN. FLUID PATH IS STERILE and NON-PYROGENIC. PRODUCT is sterilized by GAMMA IRRADIATION. CareFusion, 630-00952 Rev 01 -Infusion Set Alaris Pump Module/Medley Pump Module REF 24009-0007T, WARNI

Reason for recall

Complaints where users were unable to prime the administration set.

Distribution

Worldwide Distribution - US Nationwide including Guam International: Canada, Australia, New Zealand, EU, Middel East, South Africa, Okinawa, Saipan, Buenos Aires, Lima, Selangor, Hong Kong, Taiwan, and Budapest

Key facts

Status
Terminated
Initiation date
2018-04-20
Report date
2018-11-14
Termination date
2021-02-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source