FDA recall Z-0405-2024

Breas Medical, Inc. · Class II · device

Product

with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.

Reason for recall

Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation

Distribution

Nationwide Foreign: Canada

Key facts

Status
Ongoing
Initiation date
2023-10-13
Report date
2023-12-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
North Billerica, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0405-2024