FDA recall Z-0407-2024

Abbott Molecular, Inc. · Class II · device

Product

Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.

Reason for recall

Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2023-10-30
Report date
2023-12-06
Voluntary/Mandated
Voluntary: Firm initiated