FDA recall Z-0408-2019

Accuray Incorporated · Class II · device

Product

CyberKnife M6, Part Number 054000-004 The CyberKnife M6 Systems are indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason for recall

A robotics supplier notified Accuray of two manufacturing variations in the fastening of the in-line wrist and the casting of the wrist for the robot. The variation may result in premature failure of the component.

Distribution

worldwide distribution.

Key facts

Status
Terminated
Initiation date
2018-09-26
Report date
2018-11-14
Termination date
2021-05-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Madison, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0408-2019