FDA recall Z-0422-2024

Siemens Healthcare Diagnostics Inc · Class II · device

Product

RAPIDPOINT 500e Blood Gas System

Reason for recall

There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.

Distribution

Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Republic, Costa Rica, Cote dIvoire, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Fiji, Finland, France, French, Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mexico, Moldova, Montenegro, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestinian Territory, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Federation, Sau

Key facts

Status
Ongoing
Initiation date
2023-09-26
Report date
2023-12-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Norwood, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0422-2024